PLoS ONE (Jan 2020)

Longitudinal change of cervical artificial disc motion following replacement.

  • Jung Hyeon Moon,
  • Chun Kee Chung,
  • Chi Heon Kim,
  • Chang-Hyun Lee,
  • Sung Bae Park,
  • Won Heo

DOI
https://doi.org/10.1371/journal.pone.0228628
Journal volume & issue
Vol. 15, no. 2
p. e0228628

Abstract

Read online

We reviewed charts and radiologic studies of 30 patients operated upon by ADR with Mobi-C® in single level since 2006. All patients had healthy cervical facet joints (less than or equal to grade 1 according to grading systems for cervical facet joint degeneration) preoperatively. We assessed clinical outcomes with NDI and VAS on neck and arm over follow-up and also measured ROM at implanted segment on dynamic radiographs during follow-up. The mean follow-up period was 42.4 ± 15.9 months. We then assessed the linearity of changes in ROM at implanted segment through linear mixed model. All patients showed significantly improved clinical outcomes. ROMs at implanted segment were maintained at slightly increased levels until 24 months postoperatively (P = 0.529). However, after 24 months, ROMs at implanted segment decreased significantly until last follow-up (P = 0.001). In addition, the decreasing pattern after 24 months showed a regular regression (P = 0.001). This decline was correlated with decline of extension angle at implanted segment. Based on this regular regression, we estimated that ROMs at implanted segments would be less than 2 degrees at 10.24 years postoperatively. Even though implanted segment maintains its motion for some length of time, we could assume that an artificial disc would have limited life expectancy correlated with the decline of extension angle.