Cyborg and Bionic Systems (Jan 2023)

Three-Dimensional Collision Avoidance Method for Robot-Assisted Minimally Invasive Surgery

  • Ling Li,
  • Xiaojian Li,
  • Bo Ouyang,
  • Hangjie Mo,
  • Hongliang Ren,
  • Shanlin Yang

DOI
https://doi.org/10.34133/cbsystems.0042
Journal volume & issue
Vol. 4

Abstract

Read online

In the robot-assisted minimally invasive surgery, if a collision occurs, the robot system program could be damaged, and normal tissues could be injured. To avoid collisions during surgery, a 3-dimensional collision avoidance method is proposed in this paper. The proposed method is predicated on the design of 3 strategic vectors: the collision-with-instrument-avoidance (CI) vector, the collision-with-tissues-avoidance (CT) vector, and the constrained-control (CC) vector. The CI vector demarcates 3 specific directions to forestall collision among the surgical instruments. The CT vector, on the other hand, comprises 2 components tailored to prevent inadvertent contact between the robot-controlled instrument and nontarget tissues. Meanwhile, the CC vector is introduced to guide the endpoint of the robot-controlled instrument toward the desired position, ensuring precision in its movements, in alignment with the surgical goals. Simulation results verify the proposed collision avoidance method for robot-assisted minimally invasive surgery. The code and data are available at https://github.com/cynerelee/collision-avoidance.