Radioengineering (Apr 2008)

Broadband Measurement of Complex Permittivity Using Reflection Method and Coaxial Probes

  • R. Zajicek,
  • L. Oppl,
  • J. Vrba

Journal volume & issue
Vol. 17, no. 1
pp. 14 – 19

Abstract

Read online

This paper describes and evaluates a method for determining complex permittivity, and presents results of permittivity measurement of some substances. Complex permittivity of a phantom of biological muscle tissue, of some industrial chemicals and dielectrics is found. A non-destructive and non-invasive method based on reflection coefficient measurement of an open-ended coaxial line attaching the material under test is used. Two coaxial probes are under investigation. Vector measurement of the reflection coefficient on the interface between probes and measured samples is performed with the aid of network analyzer in the frequency range from 300 kHz to 3 GHz. Numerical modeling (FDTD) is compared with measurement. The results indicate that using the coaxial probe with dimensions of N connector the method is suitable in the frequency range approximately from 30 MHz to 1 GHz and using the probe with dimensions of SMA connector in range approximately from 30 MHz to 3 GHz.

Keywords