International Journal of Infectious Diseases (Jun 2021)

Diagnostic accuracy of LAMP versus PCR over the course of SARS-CoV-2 infection

  • Masato Inaba,
  • Yuki Higashimoto,
  • Yoko Toyama,
  • Tomoya Horiguchi,
  • Masaya Hibino,
  • Mitsunaga Iwata,
  • Kazuyoshi Imaizumi,
  • Yohei Doi

Journal volume & issue
Vol. 107
pp. 195 – 200

Abstract

Read online

Objective: Reverse transcription loop-mediated isothermal amplification (RT-LAMP) has been validated to diagnose several viral infections. However, its diagnostic accuracy in detecting SARS-CoV-2 in real-life clinical settings remains unclear. This study aimed to determine the diagnostic sensitivity and specificity of RT-LAMP compared to reverse transcription-quantitative polymerase chain reaction (RT-qPCR) over the disease course of COVID-19. Methods: A total of 124 nasopharyngeal swab samples obtained from 24 COVID-19 patients were tested by RT-LAMP and RT-qPCR. Sensitivities and specificities of RT-LAMP compared with RT-qPCR were analyzed as a function of time from onset. Results: Up to the 9th day after onset, the RT-LAMP had a positivity of 92.8%, and the sensitivity and specificity compared with RT-qPCR was 100%. However, after the 10th day after onset, the positivity of RT-LAMP decreased to less than 25%, and the concordance of positivity between the two methods was below 60%. The limit of detection of RT-LAMP was 6.7 copies/reaction. Conclusions: Until the 9th day after the onset of symptoms, RT-LAMP had the same diagnostic accuracy as RT-qPCR. These findings suggest that RT-LAMP can be used as a diagnostic tool for COVID-19 as an alternative to RT-qPCR in the acute symptomatic phase of COVID-19.