Journal of the Formosan Medical Association (Aug 2014)

Using high-performance liquid chromatography with UV detector to quantify exhaled leukotriene B4 level in nonatopic adults

  • Gwo-Hwa Wan,
  • Dah-Chin Yan,
  • Hui-Yun Tseng,
  • Jian-Tao Lee,
  • Yu-Wen Lin

DOI
https://doi.org/10.1016/j.jfma.2013.12.009
Journal volume & issue
Vol. 113, no. 8
pp. 566 – 568

Abstract

Read online

This study aimed to evaluate the feasibility of the chemical method to analyze exhaled breath condensate (EBC) leukotriene B4 (LTB4) level in humans. High-performance liquid chromatography with a UV detector was applied to quantify the inflammatory biomarker. The LTB4 concentration in the concentrated pooled EBC samples was 1.19 ng/μL, and the average LTB4 concentration of each EBC sample was 15.38 ng/μL. This analytical technique was feasible to evaluate the levels of inflammatory mediators such as LTB4 in human EBCs without any complicated sample pretreatment processes.