Sensors (Aug 2020)

Compensation System for Biomagnetic Measurements with Optically Pumped Magnetometers inside a Magnetically Shielded Room

  • Anna Jodko-Władzińska,
  • Krzysztof Wildner,
  • Tadeusz Pałko,
  • Michał Władziński

DOI
https://doi.org/10.3390/s20164563
Journal volume & issue
Vol. 20, no. 16
p. 4563

Abstract

Read online

Magnetography with superconducting quantum interference device (SQUID) sensor arrays is a well-established technique for measuring subtle magnetic fields generated by physiological phenomena in the human body. Unfortunately, the SQUID-based systems have some limitations related to the need to cool them down with liquid helium. The room-temperature alternatives for SQUIDs are optically pumped magnetometers (OPM) operating in spin exchange relaxation-free (SERF) regime, which require a very low ambient magnetic field. The most common two-layer magnetically shielded rooms (MSR) with residual magnetic field of 50 nT may not be sufficiently magnetically attenuated and additional compensation of external magnetic field is required. A cost-efficient compensation system based on square Helmholtz coils was designed and successfully used for preliminary measurements with commercially available zero-field OPM. The presented setup can reduce the static ambient magnetic field inside a magnetically shielded room, which improves the usability of OPMs by providing a proper environment for them to operate, independent of initial conditions in MSR.