Journal of Interventional Cardiology (Jan 2022)

Balloon Fracturing Valve-in-Valve: How to Do It and a Case Report of TAVR in a Rapid Deployment Prosthesis

  • Rodrigo Petersen Saadi,
  • Ana Paula Tagliari,
  • Carisi Anne Polanczyck,
  • João Carlos Ferreira Leal,
  • Eduardo Keller Saadi

DOI
https://doi.org/10.1155/2022/4368887
Journal volume & issue
Vol. 2022

Abstract

Read online

Transcatheter aortic valve replacement (TAVR) to treat degeneration of bioprosthetic heart valves (BHVs), called as valve-in-valve (ViV), is becoming a key feature since the number of BHVs requiring intervention is increasing and many patients are at high risk for a redo cardiac surgery. However, a TAVR inside a small previous cardiac valve may lead to prosthesis-patient mismatch (PPM) and not be as effective as we hoped for. An effective option to decrease the chance of PPM is to fracture the previous heart valve implanted using a high-pressure balloon. By performing a valve fracture, the inner valve ring of small BHVs can be opened up by a single fracture line, allowing subsequent implantation of a properly sized transcatheter heart valve, without increasing substantially the procedure risk. In this article, we provide a step-by-step procedure on how to safely and properly fracture a BHV and report a case of a TAVR in a degenerated rapid deployment valve.