Brazilian Journal of Pharmaceutical Sciences (Sep 2014)

Standardization method for measurement of hydroxyurea by Ultra High Efficiency Liquid Chromatography in plasma of patients with sickle cell disease

  • Darcielle Bruna Dias Elias,
  • Teresa Maria de Jesus Ponte Carvalho,
  • Janete Eliza de Sá Soares,
  • Romélia Pinheiro Gonçalves

DOI
https://doi.org/10.1590/S1984-82502014000300022
Journal volume & issue
Vol. 50, no. 3
pp. 621 – 628

Abstract

Read online

Sickle cell anemia (SCA) is a recessively inherited disease characterized by chronic hemolytic anemia, chronic inflammation, and acute episodes of hemolysis. Hydroxyurea (HU) is widely used to increase the levels of fetal hemoglobin (HbF). The objective of this study was to standardize and validate a method for the quantification of HU in human plasma by using ultra high performance liquid chromatography (UPLC) in order to determine the plasma HU levels in adult patients with SCA who had been treated with HU. We used an analytical reverse phase column (Nucleosil C18) with a mobile phase consisting of acetonitrile/water (16.7/83.3). The retention times of HU, urea, and methylurea were 6.7, 7.7, and 11.4 min, respectively. All parameters of the validation process were defined. To determine the precision and accuracy of quality controls, HU in plasma was used at concentrations of 100, 740, and 1600 µM, with methylurea as the internal standard. Linearity was assessed in the range of 50-1600 µM HU in plasma, obtaining a correlation coefficient of 0.99. The method was accurate and precise and can be used for the quantitative determination of HU for therapeutic monitoring of patients with SCA treated with HU.