BMC Research Notes (Aug 2024)

Comparison of biuret and refractometery method for serum total protein measurements in cattle and goat

  • Sileshi Mesele Mesa,
  • Yoseph Cherinet Megerssa

DOI
https://doi.org/10.1186/s13104-024-06906-5
Journal volume & issue
Vol. 17, no. 1
pp. 1 – 6

Abstract

Read online

Abstract Objectives The biuret method is frequently used to determine serum total protein. On the other hand refractometer, a quicker and less expensive option, is available to determine serum total protein. However, there is no study conducted in Ethiopia to compare serum total protein measurement in veterinary settings. Therefore, this study was conducted to compare the serum total protein concentration measurement in cattle and goats obtained by the biuret method and refractometer. Results Serum samples from 60 cattle and 60 goats were assayed by both methods and data were analyzed with a paired t-test, Pearson’s correlation, and Bland-Altman plots. There was a strong positive correlation between the total protein values determined with the refractometer and the biuret method in cattle (r = 0.93) and goats (r = 0.97). There were no significant differences (p > 0.05) in the protein values measured with the refractometer and those evaluated with the biuret method in both species. Bland-Altman plots showed that biases indicating the analytic and user error were 8.33% in both species which is below the acceptable total error (< 10%). Thus, refractometer can be used in place of biuret method since it is valid enough to measure serum total protein in cattle and goats.

Keywords