Read online

Camera resectioning is the process of finding the intrinsic and extrinsic parameters of the pinhole camera model. A mobile X-ray device (C-arm) can be represented with the pinhole camera model. The C-arm parameters can be determined with a parametrization device and the direct linear transformation (DLT) algorithm. We present the conception process of a parametrization device and further optimizations of the DLT algorithm. The evaluation of the parametrization quality is carried out with an inverse registration approach on a testing device. The seven landmarks in the testing device are compared on an X-ray image and a digitally reconstructed radiograph (DRR). The average distance of the corresponding landmarks between imaging modalities is 2.5 pixels. This deviation indicates an adequate design of the parametrization device for a 2D/3D registration in minimally invasive surgery.