Failure of volar locking plate fixation of an extraarticular distal radius fracture: A case report

  • Ozer Kagan,
  • Cao Jue

DOI
https://doi.org/10.1186/1754-9493-4-19
Journal volume & issue
Vol. 4, no. 1
p. 19

Abstract

Read online

Abstract Background Volar locking plates provide significant structural stability to the distal radius. Failure of a volar locked plating is a rarely reported complication in the literature. Case Presentation A 40 year-old, obese female patient who presented with a displaced extraarticular distal radius fracture, underwent open reduction and internal fixation of the fracture using a volar locking plate. Radiographs taken at 10 weeks postoperatively showed failure of fixation with breakage of the four distal locking screws. A hardware removal was performed at 6 months, and the patient was then lost to follow-up. She presented again at 18 months after the first surgery, with significant pain, and radiographic signs of a radial collapse and a fracture-nonunion. A total wrist fusion was performed as the method of choice at that point in time. Conclusion Volar locked plating represents the new "gold standard" of distal radius fracture fixation. However, despite the stability provided by locking plates, hardware failure may occur and lead to a cascade of complications which will ultimately require a wrist fusion, as outlined in this case report. Additional structural support by bone grafting may be needed in selected cases of volar locked plating, particularly in patients with a high risk of developing a fracture-nonunion.