Animals (Jan 2024)

Evaluation of the Audicor Acoustic Cardiography Device as a Diagnostic Tool in Horses with Mitral or Aortic Valve Insufficiency

  • Isabelle L. Piotrowski,
  • Hannah K. Junge,
  • Colin C. Schwarzwald

DOI
https://doi.org/10.3390/ani14020331
Journal volume & issue
Vol. 14, no. 2
p. 331

Abstract

Read online

Mitral and aortic valve insufficiencies have been commonly reported in horses. The objective of this study was to establish the use of acoustic cardiography (Audicor®) in horses with aortic (AI) or mitral valve insufficiency (MI). A total of 17 healthy horses, 18 horses with AI, and 28 horses with MI were prospectively included. None of the horses was in heart failure. Echocardiography and Audicor® analyses were conducted. Electromechanical activating time (EMAT), rate-corrected EMATc, left ventricular systolic time (LVST), rate-corrected LVSTc, and intensity and persistence of the third and fourth heart sound (S3, S4) were reported by Audicor®. Graphical analysis of the three-dimensional (3D) phonocardiogram served to visually detect murmurs. Audicor® snapshot variables were compared between groups using one-way ANOVA followed by Tukey’s multiple-comparisons test. The association between Audicor® snapshot variables and the corresponding echocardiographic variables was investigated by linear regression and Bland–Altman analyses. Heart murmurs were not displayed on Audicor® phonocardiograms. No significant differences were found between Audicor® variables obtained in clinically healthy horses and horses with valvular insufficiency. The Audicor® device is unable to detect heart murmurs in horses. Audicor® variables representing cardiac function are not markedly altered, and their association with corresponding echocardiographic variables is poor in horses with valvular insufficiency that are not in heart failure.

Keywords