Scientific Reports (Nov 2024)

Universal non-circular cone beam CT orbits for metal artifact reduction imaging during image-guided procedures

  • Tess Reynolds,
  • Yiqun Ma,
  • Andrew Kanawati,
  • Owen Dillon,
  • Kenzie Baer,
  • Grace Gang,
  • Joseph Stayman

DOI
https://doi.org/10.1038/s41598-024-77964-9
Journal volume & issue
Vol. 14, no. 1
pp. 1 – 13

Abstract

Read online

Abstract Innovation in image-guided procedures has been driven by advances in robotic Cone Beam Computed Tomography (CBCT) systems. A fundamental challenge for CBCT imaging is metal artifacts arising from surgical tools and implanted hardware. Here, we outline how two universal non-circular imaging orbits, optimized for metal artifact reduction, can be implemented in real-time on clinical robotic CBCT systems. Demonstrating potential clinical utility, the universal orbits were implemented during a pedicle screw cervical spine fixation and hip arthroplasty performed on a porcine and ovine cadaver respectively. In both procedures, the universal non-circular orbits noticeably reduced the metal artifacts surrounding the implanted orthopedic hardware, revealing anatomy and soft tissue obscured in current conventional CBCT imaging. This work represents a key step in clinically translating universal orbits, unlocking high quality in-room procedural verification to increase broader use of robotic CBCT systems and reduce the occurrence of secondary corrective surgeries.