Microbiology Spectrum (Jun 2023)

The Clinical Performance of the BioCode Respiratory Pathogen Panel for the Detection of Viruses and Bacteria from Nasopharyngeal Swabs

  • Xin Zhang,
  • Colleen Knoth,
  • Anh Pham,
  • Amorce Lima,
  • Rosario Dominguez,
  • Irvin Ibarra-Flores,
  • Juan C. Lopez,
  • Dominic Uy,
  • Suzane Silbert,
  • Anami Patel,
  • Michael Aye,
  • Yi-Wei Tang,
  • Jennifer Dien Bard

DOI
https://doi.org/10.1128/spectrum.04044-22
Journal volume & issue
Vol. 11, no. 3

Abstract

Read online

ABSTRACT Early detection of microbial pathogens causing respiratory tract infection plays a crucial role in clinical management. The BioCode Respiratory Pathogen Panel (BioCode RPP) utilizes reverse transcriptase PCR (RT-PCR) in combination with barcoded magnetic beads to amplify, detect, and identify respiratory pathogens. This panel qualitatively detects and identifies 14 viruses, including influenza virus A with H1 pdm09, H1, and H3 subtyping; influenza B; respiratory syncytial virus (RSV); human metapneumovirus; parainfluenza virus 1; parainfluenza virus 2; parainfluenza virus 3; parainfluenza virus 4; coronavirus (229E, NL63, OC43, and HKU1); adenovirus; and human rhinovirus/enterovirus, and 3 bacteria, including Chlamydia pneumoniae, Mycoplasma pneumoniae, and Bordetella pertussis. Reproducibility, which was assessed with contrived specimens containing 12 targets at 3 clinical sites, with 2 operators at each site for 5 days, was 99.4% for Flu A H3 and Flu B, 98.9% for RSV, and 100% for the remaining 9 targets assayed. A multicenter clinical trial evaluated the performance of the BioCode RPP with 2,647 nasopharyngeal swab specimens from 5 geographically distinct sites and revealed comparable performance between the BioCode RPP and FilmArray Respiratory Panel (FA-RP). Specifically, the positive percent agreements (PPAs) for various pathogens ranged between 80.8% and 100% compared with the FA-RP (1.7 and 2.0). Negative percent agreement ranged from 98.4% to 100% for BioCode RPP. The BioCode RPP also offers scalable automated testing capability of up to 96 specimens in a single run with total sample-to-result time under 5 h. The invalid rate of the BioCode RPP on initial testing was 1.0% (26/2,649). IMPORTANCE Early detection of microbial pathogens causing respiratory tract infection plays a crucial role in clinical management. The BioCode Respiratory Pathogen Panel (BioCode RPP) is a high-throughput test that utilizes RT-PCR in combination with barcoded magnetic beads to amplify, detect, and identify 17 respiratory pathogens, including 14 viruses and 3 bacteria. This study summarizes data generated from a multicenter clinical trial evaluating the performance of the BioCode RPP on 2,647 nasopharyngeal swab specimens from five geographically distinct sites.

Keywords