PLoS ONE (Jan 2024)
Development and performance evaluation of a solenoid valve assisted low-cost ventilator on gas exchange and respiratory mechanics in a porcine model.
Abstract
IntroductionDuring the COVID-19 pandemic, ventilator shortages necessitated the development of new, low-cost ventilator designs. The fundamental requirements of a ventilator include precise gas delivery, rapid adjustments, durability, and user-friendliness, often achieved through solenoid valves. However, few solenoid-valve assisted low-cost ventilator (LCV) designs have been published, and gas exchange evaluation during LCV testing is lacking. This study describes the development and performance evaluation of a solenoid-valve assisted low-cost ventilator (SV-LCV) in vitro and in vivo, focusing on gas exchange and respiratory mechanics.MethodsThe SV-LCV, a fully open ventilator device, was developed with comprehensive hardware and design documentation, utilizing solenoid valves for gas delivery regulation. Lung simulator testing calibrated tidal volumes at specified inspiratory and expiratory times, followed by in vivo testing in a porcine model to compare SV-LCV performance with a conventional ventilator.ResultsThe SV-LCV closely matched the control ventilator's respiratory profile and gas exchange across all test cycles. Lung simulator testing revealed direct effects of compliance and resistance changes on peak pressures and tidal volumes, with no significant changes in respiratory rate. In vivo testing demonstrated comparable gas exchange parameters between SV-LCV and conventional ventilator across all cycles. Specifically, in cycle 1, the SV-LCV showed arterial blood gas (ABG) results of pH 7.54, PCO2 34.5 mmHg, and PO2 91.7 mmHg, compared to the control ventilator's ABG of pH 7.53, PCO2 37.1 mmHg, and PO2 134 mmHg. Cycle 2 exhibited ABG results of pH 7.53, PCO2 33.6 mmHg, and PO2 84.3 mmHg for SV-LCV, and pH 7.5, PCO2 34.2 mmHg, and PO2 93.5 mmHg for the control ventilator. Similarly, cycle 3 showed ABG results of pH 7.53, PCO2 32.1 mmHg, and PO2 127 mmHg for SV-LCV, and pH 7.5, PCO2 35.5 mmHg, and PO2 91.3 mmHg for the control ventilator.ConclusionThe SV-LCV provides similar gas exchange and respiratory mechanic profiles compared to a conventional ventilator. With a streamlined design and performance akin to commercially available ventilators, the SV-LCV presents a viable, readily available, and reliable short-term solution for overcoming ventilator supply shortages during crises.