ESC Heart Failure (Oct 2024)

In vivo fluid dynamics of the Ventura interatrial shunt device in patients with heart failure

  • Michael Pfeiffer,
  • John Boehmer,
  • John Gorcsan,
  • Shunsuke Eguchi,
  • Yoshiyuki Orihara,
  • Michal Laufer Perl,
  • Neal Eigler,
  • William T. Abraham,
  • Julio Nuñez Villota,
  • Elizabeth Lee,
  • Antoni Bayés‐Genís,
  • Gil Moravsky,
  • Saibal Kar,
  • Michael R. Zile,
  • Richard Holcomb,
  • Stefan D. Anker,
  • Gregg W. Stone,
  • Josep Rodés‐Cabau,
  • JoAnn Lindenfeld,
  • Jeroen J. Bax

DOI
https://doi.org/10.1002/ehf2.14859
Journal volume & issue
Vol. 11, no. 5
pp. 2499 – 2509

Abstract

Read online

Abstract Aims Interatrial shunts are under evaluation as a treatment for heart failure (HF); however, their in vivo flow performance has not been quantitatively studied. We aimed to investigate the fluid dynamics properties of the 0.51 cm orifice diameter Ventura shunt and assess its lumen integrity with serial transesophageal echocardiography (TEE). Methods and results Computational fluid dynamics (CFD) and bench flow tests were used to establish the flow‐pressure relationship of the shunt. Open‐label patients from the RELIEVE‐HF trial underwent TEE at shunt implant and at 6 and 12 month follow‐up. Shunt effective diameter (Deff) was derived from the vena contracta, and flow was determined by the continuity equation. CFD and bench studies independently validated that the shunt's discharge coefficient was 0.88 to 0.89. The device was successfully implanted in all 97 enrolled patients; mean age was 70 ± 11 years, 97% were NYHA class III, and 51% had LVEF ≤40%. Patency was confirmed in all instances, except for one stenotic shunt at 6 months. Deff remained unchanged from baseline at 12 months (0.47 ± 0.01 cm, P = 0.376), as did the trans‐shunt mean pressure gradient (5.1 ± 3.9 mmHg, P = 0.316) and flow (1137 ± 463 mL/min, P = 0.384). TEE measured flow versus pressure closely correlated (R2 ≥ 0.98) with a fluid dynamics model. At 12 months, the pulmonary/systemic flow Qp/Qs ratio was 1.22 ± 0.12. Conclusions When implanted in patients with advanced HF, this small interatrial shunt demonstrated predictable and durable patency and performance.

Keywords