Journal of Orthopaedic Surgery and Research (Sep 2023)

Long-term clinical outcomes of the Biomet M2a-38: a retrospective review of 335 total hip arthroplasty cases

  • Caché S. Jones,
  • Dani Gaillard-Campbell,
  • Thomas P. Gross

DOI
https://doi.org/10.1186/s13018-023-04223-4
Journal volume & issue
Vol. 18, no. 1
pp. 1 – 6

Abstract

Read online

Abstract Background At the turn of the century, over one-third of total hip arthroplasties comprised metal-on-metal bearings. As this patient population and their implants age, it is crucial to understand associated late failure modes and expected long-term functional outcomes. We report the long-term results of a large metal-on-metal uncemented total hip arthroplasty system with unique design characteristics compared to others that have been reported with high failure rates. Methods We retrospectively analyze our prospective clinical database to determine overall implant survivorship and functional outcomes. Further, we compare these results to the clinical outcomes reported in orthopedic registries and in other published studies with similar metal-on-metal total hip arthroplasty cohorts. Results Implant survivorship at 10 years was 99.1% and continued to 97.6% survivorship at 20 years. Implant survivorship at 20 years did not vary significantly between sexes (Male: 98.3%, Female: 97.2%; log-rank p-value = 0.46). Mean whole blood cobalt levels were 2.6 µg/L in unilateral cases, 5.3 µg/L in bilateral patients, and 3.4 µg/L for the combined cohort. Average blood chromium levels were 1.4 µg/L in unilateral patients, 2.9 µg/L in bilateral patients, and 1.8 µg/L for group combined. We observed a 0.9% rate of failure due trunnion corrosion at a mean of 13.1 years postoperatively (10.6–15.6 years) but had no bearing wear failures. Conclusions Our 20-year implant survivorship of 97.6% with the M2a-38 bearing surpassed registry benchmarks for THA. This large-bearing (38 mm), full hemisphere coverage metal-on-metal system had no bearing wear failures, one failure of instability, one failure of fixation, and three trunnion failures, perhaps suggesting an optimum balance between stability of the joint and the trunnion.