Current Directions in Biomedical Engineering (Sep 2017)

A new device for vein localization and effect of application of disinfectant spray on its efficiency

  • Dreyer Jan,
  • Huenges Katharina,
  • Cremer Jochen,
  • Haneya Assad

DOI
https://doi.org/10.1515/cdbme-2017-0026
Journal volume & issue
Vol. 3, no. 2
pp. 127 – 130

Abstract

Read online

A functional device was developed to immediately show the localization of veins by detecting a temperature increase on the skin directly above them. Our new idea, compared to other developments, is the comparison of temperatures between a small, ideally punctiform, skin area, and a larger circularly surrounding area. This is realized by two infrared temperature sensors, one with a small field of view, and the other one with a larger field of view. The position of the vein is indicated by two laser modules, which beams cross in one spot, when the device is held in a defined distance to the skin. If the device is held over a vein, the laser spot lightens up. The device was tested in ten study participants. Cooling of the skin by disinfectant spray prior to the measurements increases the temperature gradient and thereby improves the efficiency of the device. Temperature profiles of four skin areas of each study participant were measured before and one minute after application of disinfectant spray. After application of disinfectant spray, a temperature difference of more than 0.3 K between a measuring point above a vein and points 15 mm next to this could be found in 36 out of 40 measurements (90%), compared to 26 out of 40 (65%) before disinfection. The mean temperature gradient could be increased from 0.476 K to 1.03 K (p < 0.001).