Journal of Medical Radiation Sciences (Mar 2020)

Effect of patient size on image quality in radiotherapy kV planar verification imaging: a phantom study

  • Sara Chan,
  • Eileen Giles,
  • Lyndal Newmarch,
  • Michala Short

DOI
https://doi.org/10.1002/jmrs.355
Journal volume & issue
Vol. 67, no. 1
pp. 34 – 42

Abstract

Read online

Abstract Introduction This study aimed to determine a maximal pelvic separation and waist circumference in pelvic patients to guide radiation therapists in acquiring kilovoltage (kV) planar images of acceptable quality for treatment verification. Methods A pelvic anthropomorphic phantom modified with different bolus thicknesses was imaged at various default kV exposure settings. Radiation therapists rated image quality and acceptance/rejection of these images for treatment verification. Results Sixteen radiation therapists participated in the study. Image quality was inversely proportional to phantom size. AP and lateral kV images were acceptable for treatment verification up to a waist circumference of 143 cm. Conclusions Exposure settings for kV image verification of large patients should be individualised to avoid unnecessary patient radiation dose through repeated imaging.

Keywords