Arquivos Brasileiros de Cardiologia (Mar 2007)

Ablação do flutter atrial típico: estudo prospectivo e randomizado do cateter irrigado fechado versus cateter com eletrodo distal de 8 mm Ablation of typical atrial flutter: a prospective randomized study of cooled-tip versus 8-mm-tip catheters

  • Sissy L. Melo,
  • Mauricio I. Scanavacca,
  • Francisco C. C. Darrieux,
  • Denise T. Hachul,
  • Eduardo A. Sosa

DOI
https://doi.org/10.1590/S0066-782X2007000300004
Journal volume & issue
Vol. 88, no. 3
pp. 273 – 278

Abstract

Read online

OBJETIVOS: Trabalho prospectivo, randomizado para comparar a eficácia e a segurança do cateter irrigado em relação ao cateter com eletrodo distal de 8 mm para ablação com radiofreqüência (RF) do flutter atrial. MÉTODOS: Em 52 pacientes consecutivos referidos para tratamento do flutter atrial típico, a ablação do istmo cavotricuspídeo (Ist-CT) foi realizada com cateter de irrigação fechada (n=26) ou com cateter de eletrodo distal de 8 mm (n=26). Os pulsos de RF foram aplicados ponto a ponto por 60 segundos com potência limitada a 50 w com o cateter irrigado e por controle de temperatura (60ºC, 70 w) com cateter de 8 mm. O critério de fim do procedimento foi a obtenção de bloqueio bidirecional do Ist-CT. RESULTADOS: O bloqueio Ist-CT foi obtido em 98,1% dos pacientes. O "crossover" ocorreu em quatro pacientes do grupo com cateter irrigado. Não se encontrou diferença estatística significante em relação aos parâmetros da ablação, tais como tempo total de aplicação de RF (591,1±309,0s vs 486,2±250,8s), duração do procedimento (86,4 ± 23,6 vs 78,1±22,5min) e tempo de fluoroscopia (17,0±6,7 vs 15,4±4,6min) entre os dois grupos. Durante seguimento médio de 10,6 meses, um paciente do grupo irrigado apresentou recorrência do flutter atrial típico. CONCLUSÃO: A ablação do Ist-CT resultou ser efetiva e segura para o controle do flutter atrial com ambas as técnicas empregadas (cateter com eletrodo distal de 8 mm e cateter irrigado). A complexidade técnica do cateter irrigado proporciona menor competitividade.OBJECTIVES: Both ablation catheters with closed irrigated system and 8mm tip-catheters have been shown to be more effective for typical atrial flutter radiofrequency (RF) ablation when compared to conventional 4 mm tip catheter. Considering the differences in complexity and costs of both systems, a prospective study was designed to compare the efficacy and safety of cooled-tip and 8mm-tip catheters for atrial flutter ablation. METHODS: Fifty-two consecutive patients underwent RF ablation of cavotricupsid isthmus (CTI) for the treatment of typical atrial flutter, using catheter with closed irrigation system (n=26) or 8 mm-tip catheter (n=26). The RF pulses were applied point-by-point for 60 seconds, with power limited at 50 w for the irrigated catheter and by temperature control (60ºC / 70 w) for the 8mm catheter. RESULTS: The CTI block was successfully performed in 98.1%. Four patients in the irrigated group needed to switch to the other group. There was no significant difference with regard to ablation parameters, such as total time of RF ablation (591.1±309.0s vs 486.2±250.8s), total procedure duration (86.4±23.6 vs 78.1±22.5min) and time of fluoroscopy (17.0±6.7 vs 15.4±4.6min). During follow-up of 10.6 months in average, one patient in the irrigated group had recurrence of typical atrial flutter. CONCLUSION: Efficacy and safety of CTI ablation was comparable between both techniques (irrigated catheter and 8mm-tip catheter). The complexity of irrigated catheter makes it less competitive.

Keywords