European Journal of Radiology Open (Dec 2023)
Reproducibility of quantitative ADC, T1, and T2 measurement on the cerebral cortex: Utility of whole brain echo-planar DWI with compressed SENSE (EPICS-DWI): A pilot study
Abstract
Purpose: To assess the reproducibility of ADC, T1, T2, and proton density (PD) measurements on the cortex across the entire brain using high-resolution pseudo-3D diffusion-weighted imaging using echo-planar imaging with compressed SENSE (EPICS-DWI) and 3D quantification with an interleaved Look-Locker acquisition sequence with T2 preparation pulse (3D-QALAS) in normal healthy adults. Methods: Twelve healthy participants (median age, 33 years; range, 28–51 years) were recruited to evaluate the reproducibility of whole-brain EPICS-DWI and synthetic MRI. EPICS-DWI utilizes a compressed SENSE reconstruction framework while maintaining the EPI sampling pattern. The 3D-QALAS sequence is based on multi-acquisition 3D gradient echo, with five acquisitions equally spaced in time, interleaved with a T2 preparation pulse and an inversion pulse. EPICS-DWI (b values, 0 and 1000 s/mm2) and 3D-QALAS sequence with identical voxel size on a 3.0-T MR system were performed twice (for test-retest scan). Intraclass correlation coefficients (ICCs) for ADC, T1, T2, and PD for all parcellated volume of interest (VOI) per subject on scan-rescan tests were calculated to assess reproducibility. Bland-Altman plots were used to investigate discrepancies in ADCs, T1s, T2s, and PDs obtained from the two MR scans. Results: The ICC of ADCs was 0.785, indicating “good” reproducibility. The ICCs of T1s, T2s, and PDs were 0.986, 0.978, and 0.968, indicating “excellent” reproducibility. Conclusion: The combination of EPICS-DWI and 3D-QALAS sequences with identical voxel size could reproducible ADC, T1, T2, and PD measurements for the cortex across the entire brain in healthy adults.