Journal of Clinical Medicine (Feb 2023)

Anatomical Changes after Transcatheter Edge-to-Edge Repair in Functional MR According to MitraClip Generation

  • Alberto Alperi,
  • Pablo Avanzas,
  • Javier Martinez,
  • Antonio Adeba,
  • Iria Silva,
  • Victor Leon,
  • Paula Antuna,
  • Daniel Hernández-Vaquero,
  • Noemi Barja,
  • Félix Fernández,
  • Cesar Moris,
  • Isaac Pascual

DOI
https://doi.org/10.3390/jcm12041486
Journal volume & issue
Vol. 12, no. 4
p. 1486

Abstract

Read online

Background: The use of mitral transcatheter edge-to-edge repair (TEER) is rapidly increasing. Anatomical changes have been described after TEER with the MitraClip system in patients with functional mitral regurgitation (MR), although no study has yet evaluated such anatomical impacts in patients treated with the G4 MitraClip generation. Methods: This research constituted a prospective, single-center, observational study including consecutive patients with functional MR. Mitral three-dimensional images were obtained transesophageally with echocardiography before and immediately after TEER. Patients receiving the late-generation (G4) system were compared to those receiving early-generation systems. Results: A total of 116 functional MR patients were evaluated, and 40 (34.5%) and 76 (65.5%) received a late-generation (G4) or early-generation device system, respectively. The baseline clinical and echocardiographic features were well-balanced between the groups. Overall, there was a significant reduction in mitral annular size after the intervention, and greater reductions in the anteroposterior diameter (4 mm vs. 3.54 mm, p = 0.03), annular perimeter (11.07 mm vs. 5.29 mm for 3D-perimeter, p = 0.001), and annular area (1.29 cm2 vs. 1.03 cm2, p = 0.002) were found for patients receiving the late G4 device generation compared to the early-generation systems. Conclusions: In patients with functional MR, we observed significant changes in mitral valve anatomy with a reduction in anteroposterior diameter, valve perimeter, and area. In our cohort, the extent of those changes was greater with the use of the new-generation G4 MitraClip system compared to prior device generations.

Keywords