Data in Brief (Sep 2016)
Absolute nutrient concentration measurements in cell culture media: 1H q-NMR spectra and data to compare the efficiency of pH-controlled protein precipitation versus CPMG or post-processing filtering approaches
Abstract
The NMR spectra and data reported in this article refer to the research article titled “A simple and accurate protocol for absolute polar metabolite quantification in cell cultures using q-NMR” [1]. We provide the 1H q-NMR spectra of cell culture media (DMEM) after removal of serum proteins, which show the different efficiency of various precipitating solvents, the solvent/DMEM ratios, and pH of the solution. We compare the data of the absolute nutrient concentrations, measured by PULCON external standard method, before and after precipitation of serum proteins and those obtained using CPMG (Carr-Purcell-Meiboom-Gill) sequence or applying post-processing filtering algorithms to remove, from the 1H q-NMR spectra, the proteins signal contribution. For each of these approaches, the percent error in the absolute value of every measurement for all the nutrients is also plotted as accuracy assessment. Keywords: 1H NMR, pH-controlled serum removal, PULCON, Accuracy, CPMG, Deconvolution