Photoacoustics (Feb 2023)

Multi-mechanism collaboration enhanced photoacoustic analyzer for trace H2S detection

  • Min Guo,
  • Xinyu Zhao,
  • Ke Chen,
  • Dongyu Cui,
  • Guangyin Zhang,
  • Chenxi Li,
  • Zhenfeng Gong,
  • Qingxu Yu

Journal volume & issue
Vol. 29
p. 100449

Abstract

Read online

To realize the real-time highly sensitive detection of SF6 decomposition product H2S, a multi-mechanism collaboration enhancement photoacoustic spectroscopy analyzer (MCEPA) based on acoustic resonance enhancement, cantilever enhancement and excitation light enhancement is proposed. An SF6 background gas-induced photoacoustic cell (PAC) was used for acoustic resonance (AR) enhancement of the photoacoustic signals. A fiber-optic acoustic sensor based on a silicon cantilever is optimized and fabricated. The narrow-band acoustic signal enhancement based on cantilever mechanical resonance (MR) is realized in the optimal working frequency band of the PAC. A fiber-coupled DFB cascaded an Erbium-doped fiber amplifier (EDFA) realized the light power enhancement (LPE) of the photoacoustic signals excitation source. Experimental results show that the MR of the fiber-optic silicon cantilever acoustic sensor (FSCAS) is matched with the AR of the PAC and combined with the LPE, which realizes the multi-mechanism collaboration enhancement of weak photoacoustic signals. The Allan-Werle deviation evaluation showed that the minimum detection limit of H2S in the SF6 background is 10.96 ppb when the average time is 200 s. Benefiting from the all-optimization of photoacoustic excitation and detection, the MCEPA has near-field high-sensitivity gas detection capability immune to electromagnetic interference.

Keywords