BMC Medical Imaging (Nov 2023)

Repeatability and agreement of AOCT-1000 M, RTVue XR and IOL master 500 in measuring corneal thickness mapping and axial length applying principle of optical coherence tomography

  • Hailong Ni,
  • Suzhong Xu,
  • Li Tian,
  • Jieli Mao,
  • Jing Li,
  • Na Lin,
  • Peike Hu,
  • Zhiyi Wu,
  • Xiang Chen,
  • Zhishu Bao,
  • Jingwei Zheng,
  • Peihua Yan,
  • Ruzhi Deng

DOI
https://doi.org/10.1186/s12880-023-01147-7
Journal volume & issue
Vol. 23, no. 1
pp. 1 – 11

Abstract

Read online

Abstract Purpose To evaluate the repeatability and agreement of Fourier-domain optical coherence tomography (AOCT-1000 M and RTVue XR) and partial coherence interferometry biometer (IOL Master 500) in measuring corneal thickness mapping and axial length respectively. Methods Corneal thickness was measured by AOCT-1000 M and RTVue XR. Axial lengths were measured by AOCT-1000 M and IOL Master 500. The repeatability and agreement of corneal thickness and axial length were calculated in two groups of devices. The intraclass correlation coefficient (ICC) was used to verify the repeatability of the device. The 95% confidence interval of the difference compared to the set cut-off value was used to verify the agreement between the two devices. Results A total of 60 subjects with 58 eyes were included. The central corneal thickness measured by AOCT-1000 M and RTVue XR were 504.46 ± 42.53 μm and 504.43 ± 42.89 μm respectively. The average difference between groups was 0.03 ± 4.58 μm, and the 95% confidence interval was (-1.17, 1.24), which was far less than the set threshold value of 15 μm (P < 0.001). Both RTVue XR and AOCT-1000 M had very good ICC values of central corneal thickness (0.998 and 0.994, respectively). The average axial lengths measured by AOCT-1000 M and IOL Master 500 were 24.28 ± 1.25 mm and 24.29 ± 1.26 mm respectively and the 95% confidence interval was (-0.02, 0.01), which was less than the set threshold value of 0.15 mm (P < 0.001). The ICC for both devices were 1.000. Conclusion Good repeatability and agreement were seen in measurements of central corneal thickness and axial length by AOCT-1000 M.

Keywords