Determinação simultânea de As, Cd e Pb em amostras de água purificada para hemodiálise por espectrometria de absorção atômica com forno de grafite, após otimização multivariada baseada no uso de planejamento experimental Simultaneous determination of arsenic, cadmium and lead by GF AAS in purified water samples for hemodialysis after multivariate optimization based on factorial design

  • Lisia Maria G. dos Santos,
  • Jaylei M. Gonçalves,
  • Silvana do Couto Jacob

DOI
https://doi.org/10.1590/S0100-40422008000500006
Journal volume & issue
Vol. 31, no. 5
pp. 975 – 979

Abstract

Read online

This paper reports the development of a methodology for simultaneously determining As, Cd and Pb, employing GF AAS with polarized Zeeman-effect background correction. In order to make the procedure applicable, the influence of pyrolysis and atomization temperatures and the amount of chemical modifiers were studied. Factorial and central composite designs were used to optimize these variables. Precision and accuracy of the method were investigated using Natural Water Reference material, Nist SRM 1640. Results are in agreement with certified values at the 95% confidence limit when the Student t-test is used. This methodology was used for quality control of purified water for hemodialysis.