Ophthalmology Science (Nov 2024)

Reference Database for a Novel Binocular Visual Function Perimeter: A Randomized Clinical Trial

  • Vincent Michael Patella, OD,
  • Nevin W. El-Nimri, OD, PhD,
  • John G. Flanagan, PhD,
  • Mary K. Durbin, PhD,
  • Timothy Bossie, OD,
  • Derek Y. Ho, MD, PhD,
  • Mayra Tafreshi, MBA,
  • Michael A. Chaglasian, OD,
  • David Kasanoff, OD,
  • Satoshi Inoue, MSc,
  • Sasan Moghimi, MD,
  • Takashi Nishida, MD, PhD,
  • Murray Fingeret, OD,
  • Robert N. Weinreb, MD

Journal volume & issue
Vol. 4, no. 6
p. 100583

Abstract

Read online

Purpose: To construct a comprehensive reference database (RDB) for a novel binocular automated perimeter. Design: A four-site prospective randomized clinical trial. Subjects and Controls: Three hundred fifty-six healthy subjects without ocular conditions that might affect visual function were categorized into 7 age groups. Methods: Subjects underwent comprehensive ocular examination of both eyes before enrollment. Using the TEMPO/IMOvifa automated perimeter (Topcon Healthcare/CREWT Medical Systems), each subject completed 4 binocular threshold visual field (VF) tests during a single visit: First, practice 24-2 and 10-2 tests were obtained from both eyes. Next, study 24-2 and 10-2 tests were obtained from both eyes. Test order of each sequence was randomized, and the tests were conducted under standard automated perimetry testing conditions: Goldmann stimulus size III, 3183 cd/m2 maximum stimulus intensity, and background intensity of 10 cd/m2, using AIZE-Rapid test strategy. Standard VF reliability indices were assessed. For each subject, 24-2 and 10-2 test results from 1 randomly selected eye were analyzed. Main Outcome Measures: Perimetric threshold sensitivity and reference limits for each test analysis parameter. Results: The ages of the study cohort were widely distributed, with a mean age (standard deviation [SD]) of 52.3 (18.5) years. Sex assignment was 44.0% male and 56.0% female. The majority of subjects self-identified as White (67.4%), followed by Black or African American (13.5%) and Asian (8.7%), with 14.6% self-identified as Hispanic or Latino ethnicity. Mean sensitivity (SD) was 29.1 (1.3) decibels (dB) for the 24-2 and 32.4 (1.0) dB for the 10-2 test. For the 24-2 and 10-2, mean sensitivity (SD) age-related changes averaged −0.06 (0.01) dB and −0.05 (0.01) dB per year, respectively. The normal range of pointwise threshold sensitivity increased with eccentricity and showed asymmetry around the mean, particularly notable in the 24-2 test. Mean (SD) binocular test duration was 3.18 (0.38) minutes (1 minute 35 seconds per eye) for the 24-2 test and 3.58 (0.43) minutes (1 minute 47 seconds per eye) for the 10-2 test. Conclusions: An RDB for the TEMPO/IMOvifa perimeter was established, highlighting the significance of considering both age and stimulus eccentricity in interpreting threshold VF test results. Financial Disclosure(s): Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.

Keywords