Scientific Reports (May 2024)

A simple and highly sensitive LC–MS workflow for characterization and quantification of ADC cleavable payloads

  • Shi Ya Mak,
  • Shuwen Chen,
  • Wey Jia Fong,
  • Andre Choo,
  • Ying Swan Ho

DOI
https://doi.org/10.1038/s41598-024-61522-4
Journal volume & issue
Vol. 14, no. 1
pp. 1 – 12

Abstract

Read online

Abstract Antibody–drug conjugates (ADC) payloads are cleavable drugs that act as the warhead to exert an ADC’s cytotoxic effects on cancer cells intracellularly. A simple and highly sensitive workflow is developed and validated for the simultaneous quantification of six ADC payloads, namely SN-38, MTX, DXd, MMAE, MMAF and Calicheamicin (CM). The workflow consists of a short and simple sample extraction using a methanol-ethanol mixture, followed by a fast liquid chromatography tandem mass spectrometry (LC–MS/MS) analysis. The results showed that well-validated linear response ranges of 0.4–100 nM for SN38, MTX and DXd, 0.04–100 nM for MMAE and MMAF, 0.4–1000 nM for CM were achieved in mouse serum. Recoveries for all six payloads at three different concentrations (low, medium and high) were more than 85%. An ultra-low sample volume of only 5 µL of serum is required due to the high sensitivity of the method. This validated method was successfully applied to a pharmacokinetic study to quantify MMAE in mouse serum samples.