Current Directions in Biomedical Engineering (Sep 2016)
Left cardiac atrioventricular delay and inter-ventricular delay in cardiac resynchronization therapy responder and non-responder
Abstract
Cardiac resynchronization therapy is an established therapy for heart failure patients. The aim of the study was to evaluate electrical left cardiac atrioventricular delay and interventricular desynchronization in sinus rhythm cardiac resynchronization therapy responder and non-responder. Cardiac electrical desynchronization were measured by surface ECG and focused transesophageal bipolar left atrial and left ventricular ECG before implantation of cardiac resynchronization therapy defibrillators. Preoperative electrical cardiac desynchronization was 195.7 ± 46.7 ms left cardiac atrioventricular delay and 74.8 ± 24.5 ms interventricular delay in cardiac resynchronization therapy responder. Cardiac resynchronization therapy responder New York Heart Association class improved during long term biventricular pacing. Transesophageal left cardiac atrioventricular delay and interventricular delay may be additional useful parameters to improve patient selection for cardiac resynchronization therapy.