Photoacoustics (Sep 2022)

Fully integrated photoacoustic microscopy and photoplethysmography of human in vivo

  • Joongho Ahn,
  • Jin Woo Baik,
  • Yeonggeon Kim,
  • Karam Choi,
  • Jeongwoo Park,
  • Hyojin Kim,
  • Jin Young Kim,
  • Hyung Ham Kim,
  • Sung Hyun Nam,
  • Chulhong Kim

Journal volume & issue
Vol. 27
p. 100374

Abstract

Read online

Photoacoustic microscopy (PAM) is used to visualize blood vessels and to monitor their time-dependent changes. Photoplethysmography (PPG) measures hemodynamic time-series changes such as heart rate. However, PPG’s limited visual access to the dynamic changes of blood vessels has prohibited further understanding of hemodynamics. Here, we propose a novel, fully integrated PAM and photoplethysmography (PAM-PPG) system to understand hemodynamic features in detail. Using the PAM-PPG system, we simultaneously acquire vascular images (by PAM) and changes in the blood volume (by PPG) from human fingers. Next, we determine the heart rate from changes in the PA signals, which match well with the PPG signals. These changes can be measured if the blood flow is not blocked. From the results, we believe that PAM-PPG could be a useful clinical tool in various clinical fields such as cardiology and endocrinology.