Brazilian Journal of Cardiovascular Surgery (Apr 2020)
Hemodynamic and Imaging Assessment of Transcatheter Aortic Valve Replacement with the Inovare® Proseal using Multislice Computed Tomography
Abstract
Abstract Objective: To evaluate the hemodynamic performance (i.e., gradients and paravalvular leakage [PVL]) of the new and experimental Braile Inovare® Proseal. Additionally, we aimed to assess pre and postoperatively the aortic annulus and the transcatheter prosthesis using multislice computed tomography (MSCT). Methods: Patients were selected by a multidisciplinary heart team and referred for transcatheter aortic valve replacement (TAVR). MSCT was performed before and after surgery. Measurements of the aortic valve and prosthesis were conducted and correlated with the valve gradient and residual PVL. Results: Twenty-one patients were selected for the protocol. Patients had a mean age of 79 years and 38% of them were of female sex. The mean EuroSCORE II value was 12.5%±10.8. Mean gradient was reduced from 45.8±11.04 mmHg to 5.59±2.61 mmHg and there were no instances of PVL worse than mild. There were no cases of coronary obstruction or procedural death. Circularity was present in all prostheses evaluated. Circularity indexes for the prostheses were: inflow 0.05±0.03, middle third 0.04±0.02, and outflow 0.04±0.02 (P=0.08). The mean distance between the prosthesis and the left and right coronary ostia were 14.8 mm±3.3 and 17.3 mm±3, respectively. Oversizing was appropriate with a mean of 22.14%±6%. Conclusion: Braile Inovare® Proseal transcatheter device has demonstrated low gradients with low rates of PVL. Oversizing by annular measurements was adequate. MSCT was adequate to evaluate device sizing and has demonstrated preserved expansibility and circularity in the evaluated cases.
Keywords