Iranian Journal of Physics Research (Nov 2021)

Monte Carlo Simulation and Experimental Determination of Tissue Phantom Ratio for Photon Beams delivered from Medical Linear Accelerator

  • N. M. Rasel,
  • S Purohit,
  • M. S. Rahman,
  • AKM M H Meaze,
  • M. Y. Ali

Journal volume & issue
Vol. 21, no. 3
pp. 55 – 61

Abstract

Read online

For an external radiotherapy procedure, the tissue phantom ratio (TPR20,10) is used as a quality index. This work presents an estimate of TPR20,10 using two cylindrical ionization chambers (NE2571 Farmer and PTW30013) in three high-energy photon modes (6, 10 and 15 MV) using both the Monte Carlo (MCNP) process and the experimental setup. The MCNP (version MCNP5) was used for the simulation of photon beams delivered by Varian-2300CD for the determination of TPR20,10 according to technical report series (TRS) 398. Again applying the same protocol TPR20,10 values were measured experimentally with NE2571 Farmer and PTW30013 chambers for the same medical linear accelerator (LINAC). The differences of TPR20, 10 between MCNP and experimental values were found for NE2571 Farmer chamber within 4.17 percent, 2.9 percent and 2.5 percent and similarly, these were within 3.89 percent, 2.71 percent and 1.98 percent at 6, 10 and 15 MV respectively for PTW30013. The TPR20,10 values simulated by our calculated MCNP demonstrated strong agreement with our experimental results.

Keywords