ETRI Journal (Jun 2024)

Miniaturized meander-line dual-band implantable antenna for biotelemetry applications

  • Rajagopal Kangeyan,
  • Madurakavi Karthikeyan

DOI
https://doi.org/10.4218/etrij.2023-0050
Journal volume & issue
no. 3
pp. 413 – 420

Abstract

Read online

A dual-band bio-implantable compact antenna with a meander-line structure is presented. The proposed meander-line antenna resonates at the industrial, scientific, and medical (2.4 GHz) and wireless medical telemetry (1.4 GHz) bands. The meander-line structure is selected as a radiating patch given its versatile and effective design. With a dimension of only 10mm X 10mm X 0:635mm, the designed antenna is compact. Considering a skin phantom, the proposed antenna was designed, optimized, and simulated. The Rogers RT/duroid 6010 substrate material with high dielectric constant was used to fabricate the meander-line dual-band implantable antenna, which was validated experimentally. The superstrate was made of the same material. Experiments were conducted on skin-mimicking gel. The designed meander-line antenna has a high peak gain of -21 dBi at 2.4 GHz, and its maximum specific absorption rate is compliant with IEEE safety standards.

Keywords