International Journal of Ophthalmology (Dec 2016)

Comparison of ICare and IOPen vs Goldmann applanation tonometry according to international standards 8612 in glaucoma patients

  • Milena Pahlitzsch,
  • Jeanette Brünner,
  • Johannes Gonnermann,
  • Anna-Karina B. Maier,
  • Necip Torun,
  • Eckart Bertelmann,
  • Matthias KJ. Klamann

DOI
https://doi.org/10.18240/ijo.2016.11.14
Journal volume & issue
Vol. 9, no. 11
pp. 1624 – 1628

Abstract

Read online

AIM: To compare IOPen and ICare rebound tonometry to Goldmann applanation tonometry (GAT) according to International Standards Organization (ISO) 8612 criteria. METHODS: Totally 191 eyes (n=107 individuals) were included. Criteria of ISO 8612 were fulfilled: 3 clusters of IOP, measured by GAT, were formed. The GAT results were given as mean±standard deviation. RESULTS: GAT (19.7±0.5 mm Hg) showed a significant correlation to ICare (19.8±0.5 mm Hg) (r=0.547, P<0.001) and IOPen (19.5±0.5 mm Hg) (r=0.526, P<0.001). According to ISO 8612 criteria in all 3 IOP groups the number of outliers (of the 95% limits of agreement) exceeded 5% for ICare and IOPen vs GAT: No.1 (n=68) 29.4% and 22.1%, No.2 (n=62) 35.5% and 37.1%, No.3 (n=61) 26.2% and 42.6%, respectively. CONCLUSION: The strict requirements of the ISO 8612 are not fulfilled in a glaucoma collective by ICare and IOPen at present. As long as the Goldmann tonometry is applicable it should be used first of all for reproducible IOP readings. ICare and IOPen tonometry should be considered as an alternative tool, if application of Goldmann tonometry is not possible.