Applied Sciences (Mar 2020)

Intraocular Pressure Study in Ex Vivo Pig Eyes by the Laser-Induced Cavitation Technique: Toward a Non-Contact Intraocular Pressure Sensor

  • Santiago Camacho-Lopez,
  • Carlos Andrés Zuñiga-Romero,
  • Luis Felipe Devia-Cruz,
  • Carolina Alvarez-Delgado,
  • Marcos Antonio Plata-Sanchez,
  • Leopoldo Martinez-Manuel

DOI
https://doi.org/10.3390/app10072281
Journal volume & issue
Vol. 10, no. 7
p. 2281

Abstract

Read online

Traditional applanation tonometry techniques lack the necessary accuracy and reliability for measuring the intraocular pressure (IOP), and there is still a need for a reliable technique for in vivo diagnosis. A single laser-induced cavitation bubble event was optically monitored in order to precisely measure the first collapse time of the cavitation bubble, which presents a direct dependence on the liquid pressure. This can certainly be done within the IOP range. We now extend the partial transmittance modulation (STM) technique to determine its feasibility for directly measuring the IOP by studying the nanosecond (ns) pulsed laser-induced cavitation bubble dynamics for an externally pressurized fresh ex vivo porcine eye. The results demonstrate that it is possible to monitor the IOP by detecting the light of a continuous-wave (CW) laser beam which is intensity modulated by the bubble itself. This technique currently presents a measurement resolution of about 4 mmHg in the 5 to 50 mmHg pressure range, indicating the feasibility of this approach for measuring IOP. This technique provides a direct measurement within the anterior eye chamber, avoiding common pitfalls in IOP diagnosis, such as errors due to patient movement, varying physical properties of the eye globe, or central cornea thickness (CCT) effects.

Keywords