Diagnostics (Mar 2024)

Short-Wavelength Infrared Imaging of Infected and Affected Dentin

  • Morgan Ng,
  • Yi-Ching Ho,
  • Spencer Wycoff,
  • Yihua Zhu,
  • Daniel Fried

DOI
https://doi.org/10.3390/diagnostics14070744
Journal volume & issue
Vol. 14, no. 7
p. 744

Abstract

Read online

Stains produced by bacteria or those found in blood and food byproducts accumulate in highly porous caries lesions. They can interfere with accurate diagnosis and the selective removal of carious tissue during cavity preparations. Short-wavelength infrared (SWIR) imaging studies have shown that stain molecules do not absorb light beyond 1200 nm. The objective of this study was to image affected and infected dentin atSWIR wavelengths. Sections of 3 mm thickness were cut from the extracted teeth with deep dentinal lesions. The sound (normal), affected (stained), and infected (demineralized) dentin on each section were examined with reflected light at wavelengths from 400 to 1700 nm, red and green fluorescence, and with optical coherence tomography (OCT). Microcomputed tomography (microCT) was used to measure the mineral density at each location investigated. Significant (p < 0.05) differences were observed in the reflected light intensity at 400–850 nm and for fluorescence between the sound, affected, and infected dentin. SWIR imaging did not show significant reductions in reflectivity for the affected and infected dentin. SWIR images may be valuable for monitoring the lateral spread of dentinal lesions on the occlusal surfaces of teeth.