IEEE Open Journal of Engineering in Medicine and Biology (Jan 2024)

A Modified Quadrature Birdcage Coil Incorporated With a Curved Feature for <italic>In Ovo</italic> MR Imaging

  • Chang-Hoon Choi,
  • Maximilian Bruch,
  • Suk-Min Hong,
  • Sandra Krause,
  • Carina Stegmayr,
  • Stefan Schwan,
  • Wieland A. Worthoff,
  • Jorg Felder,
  • N. Jon Shah

DOI
https://doi.org/10.1109/OJEMB.2024.3420231
Journal volume & issue
Vol. 5
pp. 534 – 541

Abstract

Read online

Goal: This study presents a novel MRI coil design approach explicitly tailored for chick embryo measurements, with the primary objective of improving sensitivity and coverage. Methods: The limitations posed by conventional birdcage coils were addressed by introducing a curvature feature into a standard coil. The performance of the modified coil was assessed using EM simulations and experimental evaluations, which were subsequently validated using a 7 T MRI scanner. A comparative analysis was conducted against a standard quadrature low-pass birdcage coil to evaluate key factors. Results: The proposed coil demonstrated improved SNR and uniformity, particularly in the proximity of the end-rings. These results were consistent with the findings obtained from the simulations. Conclusions: The use of our innovative birdcage coil design holds promise and offers practical potential for in ovo studies.