BMC Biotechnology (Sep 2022)

A new method for screening acute/chronic lymphocytic leukemia: dual-label time-resolved fluorescence immunoassay

  • Zhen Zhang,
  • Jintao Zhang,
  • Shanshan Dai,
  • Hang Xu

DOI
https://doi.org/10.1186/s12896-022-00758-2
Journal volume & issue
Vol. 22, no. 1
pp. 1 – 8

Abstract

Read online

Abstract Background Lymphocytic leukemia (LL) is a primary malignant tumor of hematopoietic tissue, which seriously affects the health of children and the elderly. The study aims to establish a new detection method for screening acute/chronic LL using time-resolved fluorescence immunoassay (TRFIA) via quantitative detection of S100 calcium binding protein A8 (S100A8) and leucine-rich alpha-2-glycoprotein 1 (LRG1) in serum. Methods Here a sandwich TRFIA was optimized and established: Anti-S100A8/LRG1 caputre antibodies immobilized on 96-well plates captured S100A8/LRG1, and then banded together with the anti-S100A8/LRG1 detection antibodies labeled with Europium(III) (Eu3+)/samarium(III) (Sm3+) chelates. Finally time resolved fluorometry measured the fluorescence intensity. Results The sensitivity of S100A8 was 1.15 ng/mL(LogY = 3.4027 + 0.4091 × LogX, R 2 = 0.9828, P < 0.001, dynamic range: 2.1–10,000 ng/mL), and 3.2 ng/mL for LRG1 (LogY = 3.3009 + 0.4082 × LogX, R 2 = 0.9748, P < 0.001, dynamic range: 4.0–10,000 ng/mL). The intra-assay and inter-assay CVs were low, ranging from 5.75% to 8.23% for S100A8 and 5.30% to 9.45% for LRG1 with high specificity and affinity in serum samples. Bland–Altman plots indicated TRFIA and ELISA kits have good agreement in clinical serum samples. Additionally, the cutoff values for S100A8 and LRG1 were 1849.18 ng/mL and 588.08 ng/mL, respectively. Conclusion The present TRFIA method could be used for the quantitative detection of S100A8 and LRG1 in serum, and it has high sensitivity, accuracy and specificity. Clinically, this TRFIA method could be suitable for screening of LL via the quantitative detection of S100A8 and LRG1.

Keywords