Nuclear Engineering and Technology (Apr 2020)

Development of a position sensitive CsI(Tl) crystal array

  • Guo-Zhu Shi,
  • Ruo-Fu Chen,
  • Kun Chen,
  • Ai-Hua Shen,
  • Xiu-Ling Zhang,
  • Jin-Da Chen,
  • Cheng-Ming Du,
  • Zheng-Guo Hu,
  • Guang-Wei Fan

Journal volume & issue
Vol. 52, no. 4
pp. 835 – 840

Abstract

Read online

A position-sensitive CsI(Tl) crystal array coupled with the multi-anode position sensitive photomultiplier tube (PS-PMT), Hamamatsu H8500C, has been developed at the Institute of Modern Physics. An effective, fast, and economical readout circuit based on discretized positioning circuit (DPC) bridge was designed for the 64-channel multi-anode flat panel PSPMT. The horizontal and vertical position resolutions are 0.58 mm and 0.63 mm respectively for the 1.0 × 1.0 × 5.0 mm3 CsI(Tl) array, and the horizontal and vertical position resolutions are 0.86 mm and 0.80 mm respectively for the 2.0 × 2.0 × 10.0 mm3 CsI(Tl) array. These results show that the CsI(Tl) crystal array with low cost could be applied in the fields of medical imaging and high-resolution gamma camera. Keywords: CsI(Tl), Crystal array, DPC, Discretized positioning circuit, IQSP518 system