Scanning Electron Microscopic Evaluation of the Internal Fit Accuracy of 3D-Printed Biphasic Calcium Phosphate Block: An Ex Vivo Pilot Study

  • Su-Hee Jeon,
  • Young Woo Song,
  • Jae-Kook Cha,
  • Jeong-Won Paik,
  • Sang-Sun Han,
  • Seong-Ho Choi

DOI
https://doi.org/10.3390/ma14061557
Journal volume & issue
Vol. 14, no. 6
p. 1557

Abstract

Read online

The aim of this study was to assess the internal fit accuracy of a three-dimensional (3D)-printed biphasic calcium phosphate (BCP) block compared with a 3D-milled poly methyl methacrylate (PMMA) block by scanning electron microscope (SEM) analysis. In a total of 20 porcine rib bones, two different types of defects having two adjacent walls and a floor were produced: a defect with a flat floor (flat defect; N = 10) and a defect with a concave floor (curved defect; N = 10). Each defect was grafted with either the 3D-printed BCP block or the 3D-milled PMMA block fabricated following the computer aided design. The defects were then cut cross-sectionally and evaluated under the SEM. The extents of internal contact and gap were measured and statistically analyzed (p p p > 0.05). The internal fit accuracy of the 3D-printed BCP block was reliable in both the flat and curved defects when compared with the accuracy of the 3D-milled PMMA block.