Comparative analysis of the mechanical limits of resistance in implant/abutment set of a new implant design: An in vitro study.

  • Marco Aurélio Bianchini,
  • Nilton De Bortoli Junior,
  • Berenice Anina Dedavid,
  • Piedad N De Aza,
  • Sergio Alexandre Gehrke

DOI
https://doi.org/10.1371/journal.pone.0280684
Journal volume & issue
Vol. 18, no. 1
p. e0280684

Abstract

Read online

ObjectiveThe aim of the present in vitro study was to evaluate the resistance on quasi-static forces and in the fatigue mechanical cycling of a new implant design compared to two other conventional implant designs.Materials and methodsEighty-eight implants with their respective abutments were tested and distributed into four groups (n = 22 per group): Morse taper connection implant (MT group), conventional external hexagon implant (EH con group), new Collo implant of external hexagon with the smooth portion out of the bone insertion (EH out group), and new Collo implant of external hexagon with the implant platform inserted to the bone level (EH bl group). All the sets were subjected to quasi-static loading in a universal testing machine, and we measured the maximum resistance force supported by each sample. Another 12 samples from each group were submitted to the cyclic fatigue test at 4 intensities of forces (n = 3 per force): 80%, 60%, 40%, and 20%. The number of cycles supported by each sample at each force intensity was evaluated.ResultsThe three groups of implants with external hexagon connection had similar maximum strength values of the sets (p > 0.05). Meanwhile, samples from the MT group showed the highest resistance values in comparison to the other three groups (p ConclusionsWithin the limitations of the present in vitro study, the results showed that the new Collo implant performs better when installed at bone level.