Comparison of a Blood Self-Collection System with Routine Phlebotomy for SARS-CoV-2 Antibody Testing

  • Douglas Wixted,
  • Coralei E. Neighbors,
  • Carl F. Pieper,
  • Angie Wu,
  • Carla Kingsbury,
  • Heidi Register,
  • Elizabeth Petzold,
  • L. Kristin Newby,
  • Christopher W. Woods

DOI
https://doi.org/10.3390/diagnostics12081857
Journal volume & issue
Vol. 12, no. 8
p. 1857

Abstract

Read online

The Coronavirus Disease 2019 (COVID-19) pandemic forced researchers to reconsider in-person assessments due to transmission risk. We conducted a pilot study to evaluate the feasibility of using the Tasso-SST (Tasso, Inc, Seattle, Washington) device for blood self-collection for use in SARS-CoV-2 antibody testing in an ongoing COVID-19 prevalence and immunity research study. 100 participants were recruited between January and March 2021 from a previously identified sub-cohort of the Cabarrus County COVID-19 Prevalence and Immunity (C3PI) Study who were under-going bimonthly COVID-19 antibody testing. Participants were given a Tasso-SST kit and asked to self-collect blood during a scheduled visit where trained laboratory personnel performed routine phlebotomy. All participants completed an after-visit survey about their experience. Overall, 70.0% of participants were able to collect an adequate sample for testing using the device. Among those with an adequate sample, there was a high concordance in results between the Tasso-SST and phlebotomy blood collection methods (Cohen’s kappa coefficient = 0.88, Interclass correlation coefficient 0.98 [0.97, 0.99], p < 0.0001). The device received a high-level (90.0%) of acceptance among all participants. Overall, the Tasso-SST could prove to be a valuable tool for seroprevalence testing. However, future studies in larger, diverse populations over longer periods may provide a better understanding of device usability and acceptance among older participants and those with comorbidities in various use scenarios.

Keywords