Scientific Reports (Jun 2022)
Ex-vivo study on the surface quality of corneal lenticule and stroma after low energy femtosecond laser lenticule extraction
Abstract
Abstract This study aimed to assess the surface quality of cap, stroma and lenticular surfaces created using low-energy femtosecond laser lenticule extraction (Ziemer FEMTO LDV Z8). Twenty-four porcine eyes were divided into four groups (n = 6 each): two with optimal laser power (32%) with posterior curvature equivalent to a spherical correction of -2D and -5D, respectively and the other two with high power (64%) with spherical correction of -2D and -5D respectively. Samples were analysed using scanning electron microscopy (SEM). Surface morphology was evaluated using a standard scoring system; surface relief, surface regularity, extent and position of irregularities were graded by four independent clinicians. Eyes with 32% power and -2D correction had significantly less size of the irregular area than those with -5D; however, no significant difference was found between the two groups with 64% power. When comparing eyes with -2D correction, the size of the irregular area was lesser with 32% power. Surface relief was lesser with -5D correction with 32% power than 64% power. Low-energy femtosecond laser lenticule extraction (Ziemer FEMTO LDV Z8) produces good surface quality results. There is a tendency for smoother surface stromal quality with lower power settings than a higher power.