Diagnostics (Nov 2022)

Changing the Patient’s Position: Pitfalls and Benefits for Radiation Dose and Image Quality of Computed Tomography in Polytrauma

  • Jessica Graef,
  • Keno K. Bressem,
  • Patrick Asbach,
  • Bernd Hamm,
  • Stefan M. Niehues

DOI
https://doi.org/10.3390/diagnostics12112661
Journal volume & issue
Vol. 12, no. 11
p. 2661

Abstract

Read online

For computed tomography (CT), representing the diagnostic standard for trauma patients, image quality is essential. The positioning of the patient’s arms next to the abdomen causes artifacts and is also considered to increase radiation exposure. The aim of this study was to evaluate the effect of various positionings during different CT examination steps on the extent of artifacts as well as radiation dose using iterative reconstruction (IR). 354 trauma-CTs were analyzed retrospectively. All datasets were reconstructed using IR and three different examination protocols were applied. Arm elevation led to a significant improvement of the image quality across all examination protocols (p p = 0.123). Only elevation during scout acquisition resulted in the reduction of radiation exposure (p < 0.001). To receive high-quality CT images, patients should be placed with elevated arms for the trunk scan, as artifacts remain even with the IR. Arm repositioning during the examination itself had no effect on the applied radiation dose because its modulation refers to the initial scout obtained. In order to achieve a dose effect by different positioning, a two-scout protocol (dual scout) should be used.

Keywords