Journal of Ophthalmology (Jan 2017)

Bruch’s Membrane Opening Minimum Rim Width Measurement with SD-OCT: A Method to Correct for the Opening Size of Bruch’s Membrane

  • Robert Kromer,
  • Martin Stephan Spitzer

DOI
https://doi.org/10.1155/2017/8963267
Journal volume & issue
Vol. 2017

Abstract

Read online

A precise evaluation of the retinal nerve fiber layer thickness (RNFLT) is key for diagnosing and monitoring glaucoma. The Bruch’s membrane opening minimum rim width (BMO-MRW) has been proposed as a reproducible assessment of the optic nerve. The BMO-MRW measures the minimum distance from the BMO to the internal limiting membrane. We propose an approach to correct the BMO-MRW using the BMO size for increased accuracy in interindividual comparisons in future studies. Eighty-one healthy patients received SPECTRALIS spectral domain optical coherence tomography measurements for the peripapillary RNFLT and BMO-MRW. We calculated a BMO size-corrected BMO-MRW using the mean BMO size of our cohort. BMO size was defined using the manufacturer-provided BMO area and manually measured BMO perimeter. We observed that the BMO-MRW correlated highly with the perimeter (r=−0.553, p<0.0001) and the area of the BMO (r=−0.546, p<0.0001). Using these parameters, we provided a corrected BMO size-adjusted BMO-MRW which was better correlated with the RNFLT compared to the noncorrected one (z=−3.3495, p=0.0004). We demonstrated the dependency of the BMO-MRW on ONH size. Furthermore, we showed the superiority of the corrected BMO-MRW using either the manually measured optic nerve head perimeter or the automatically provided ONH for future studies.