PLoS ONE (Jan 2019)

Validity of six consumer-level activity monitors for measuring steps in patients with chronic heart failure.

  • Tomas Vetrovsky,
  • Michal Siranec,
  • Jitka Marencakova,
  • James J Tufano,
  • Vaclav Capek,
  • Vaclav Bunc,
  • Jan Belohlavek

DOI
https://doi.org/10.1371/journal.pone.0222569
Journal volume & issue
Vol. 14, no. 9
p. e0222569

Abstract

Read online

IntroductionAlthough numerous activity trackers have been validated in healthy populations, validation is lacking in chronic heart failure patients who normally walk at a slower pace, making it difficult for researchers and clinicians to implement activity monitors during physical activity interventions.MethodsSix consumer-level activity monitors were validated in a 3-day field study in patients with chronic heart failure and healthy individuals under free living conditions. Furthermore, the same devices were evaluated in a lab-based study during treadmill walking at speeds of 2.4, 3.0, 3.6, and 4.2 km·h-1. Concordance correlation coefficients (CCC) were used to evaluate the agreement between the activity monitors and the criterion, and mean absolute percentage errors (MAPE) were calculated to assess differences between each device and the criterion (MAPE ResultsIn the field study of healthy individuals, all but one of the activity monitors showed a substantial correlation (CCC ≥0.95) with the criterion device and MAPE ConclusionsEven though none of the tested activity monitors fall within arbitrary thresholds for validity, most of them perform reasonably well enough to be useful tools that clinicians can use to simply motivate chronic heart failure patients to walk more.