Micromachines (Apr 2023)

A Thermopile-Based Gas Flow Sensor with High Sensitivity for Noninvasive Respiration Monitoring

  • Zemin Liu,
  • Chenchen Zhang,
  • Xuefeng Ding,
  • Yue Ni,
  • Na Zhou,
  • Yanhong Wang,
  • Haiyang Mao

DOI
https://doi.org/10.3390/mi14050910
Journal volume & issue
Vol. 14, no. 5
p. 910

Abstract

Read online

In this work, a N/P polySi thermopile-based gas flow device is presented, in which a microheater distributed in a comb-shaped structure is embedded around hot junctions of thermocouples. The unique design of the thermopile and the microheater effectively enhances performance of the gas flow sensor leading to a high sensitivity (around 6.6 μV/(sccm)/mW, without amplification), fast response (around 35 ms), high accuracy (around 0.95%), and mood long-term stability. In addition, the sensor has the advantages of easy production and compact size. With such characteristics, the sensor is further used in real-time respiration monitoring. It allows detailed and convenient collection of respiration rhythm waveform with sufficient resolution. Information such as respiration periods and amplitudes can be further extracted to predict and alert of potential apnea and other abnormal status. It is expected that such a novel sensor could provide a new approach for respiration monitoring related noninvasive healthcare systems in the future.

Keywords