Advanced Pharmaceutical Bulletin (Aug 2013)

Development and Application of an HPLC Method for Erlotinib Protein Binding Studies

  • Soheila Bolandnazar,
  • Adeleh Divsalar,
  • Hadi Valizadeh,
  • Arash Khodaei,
  • Parvin Zakeri-Milani

DOI
https://doi.org/10.5681/apb.2013.047
Journal volume & issue
Vol. 3, no. 2
pp. 289 – 293

Abstract

Read online

Purpose: The aim of the present study was to develop a simple and rapid reversed-phase high performance liquid chromatographic method with UV detection for erlotinib hydrochloride quantification, which is applicable for protein binding studies. Methods: Ultrafilteration method was used for protein binding study of erlotinib hydrochloride. For sample analysis a simple and rapid reversed-phase high performance liquid chromatographic method with UV detection at 332 nm was developed. The mobile phase was a mixture of methanol, acetonitril and potassium dihydrogen phosphate buffer (15:45:40 %v/v) set at flow rate of 1.3 ml/min. Results: The run time for erlotinib hydrochloride was approximately 6 minutes. The calibration curve was linear over the range of 320-20000 ng/ml with acceptable intra- and inter-day precision and accuracy. The intra-day and inter-day precisions were less than 10% and the accuracies of intra and inter-day assays were within the range of 97.20-104.83% and 98.8-102.2% respectively. Conclusion: Based on the obtained results, a simple, accurate and precise reversed-phase isocratic HPLC method with UV detection has been optimized and validated for the determination of erlotinib hydrochloride in biological samples.

Keywords