Journal of Mass Spectrometry and Advances in the Clinical Lab (Nov 2021)

Rapid identification of plasmalogen molecular species using targeted multiplexed selected reaction monitoring mass spectrometry

  • Abul Kalam Azad,
  • Hironori Kobayashi,
  • Abdullah Md. Sheikh,
  • Harumi Osago,
  • Hiromichi Sakai,
  • Md. Ahsanul Haque,
  • Shozo Yano,
  • Atsushi Nagai

Journal volume & issue
Vol. 22
pp. 26 – 33

Abstract

Read online

Plasmalogens (Pls) levels are reported to be altered in several neurological and metabolic diseases. Identification of sn-1 fatty alcohols and sn-2 fatty acids of different Pls species is necessary to determine the roles and mechanisms of action of Pls in different diseases. Previously, full-scan tandem mass spectrometry (MS/MS) was used for this purpose but is not effective for low-abundance Pls species. Recently, multiplexed selected reaction monitoring MS (SRM/MS) was found to be more selective and sensitive than conventional full-scan MS/MS for the identification of low-abundance compounds. In the present study, we developed a liquid chromatography (LC)-targeted multiplexed SRM/MS system for the identification and quantification of different Pls choline (Pls-PC) and Pls ethanolamine (Pls-PE) species. We determined five precursor-product ion transitions to identify sn-1 and sn-2 fragments of each Pls species. Consequently, sn-1 and sn-2 fatty acyl chains of 22 Pls-PC and 55 Pls-PE species were identified in mouse brain samples. Among them, some species had C20:0 and C20:1 fatty alcohols at the sn-1 position. For quantification of Pls species in mouse brain samples, a single SRM transition was employed. Thus, our results suggest that the LC-targeted multiplexed SRM/MS system is very sensitive for the identification and quantification of low-abundance lipids such as Pls, and is thus expected to make a significant contribution to basic and clinical research in this field in the future.

Keywords