European Journal of Radiology Open (Jan 2020)

Towards establishment of diagnostic reference levels based on clinical indication in the state of Qatar

  • Huda AlNaemi,
  • Virginia Tsapaki,
  • Ahmed J. Omar,
  • Maryam AlKuwari,
  • Amal AlObadli,
  • Shady Alkhazzam,
  • Antar Aly,
  • Mohammad Hassan Kharita

Journal volume & issue
Vol. 7
p. 100282

Abstract

Read online

Objectives: The objectives of this study were to: 1) evaluate patient radiation exposure in CT and 2) establish CT Diagnostic Reference Levels (DRL)s based on clinical indication (CI) in Qatar. Materials and Methods: Patient data for 13 CIs were collected using specially designed collection forms from the dose management software (DMS) of Hamad Medical Corporation (HMC), the main Qatar healthcare provider. The methodology described in the International Commission on Radiological Protection (ICRP) Report 135 was followed to establish national clinical DRLs in terms of Volumetric Computed Tomography Dose Index (CTDIvol) and total Dose Length Product (DLPt). Effective dose (Ef) was estimated by DMS using DLPt and appropriate conversion factors and was analyzed for comparison purposes. Results: Data were retrospectively collected for 896 adult patients undergoing CT examinations in 4 hospitals and 7 CT scanners. CT for Diffuse infiltrative lung disease imparted the lowest radiation in terms of CTDIvol (5 mGy), DLPt (181 mGy.cm) and Ef (3.6 mSv). Total body CT for severe trauma imparted the highest DLPt (3137 mGy.cm) and Ef (38.6 mSv) of all CIs with a CTDIvol of 15 mGy. Rounded Third quartile CTDIvol and DLPt values were defined as the Qatar CT clinical DRLs. Comparison was limited due to sparse international literature. When this was possible data were lower or comparable with other studies. Conclusions: This is the first study reporting national clinical DRLs in Asia and second one internationally after UK. For accurate comparison between studies, systemized CI nomenclature must be followed by researchers.

Keywords